Thermometer Corrections CHM 254
and 265 Labs
Throughout
the course, you will need to measure the temperature of a liquid or a vapor in
various experiments. Although the
thermometers should provide accurate measurements of temperature, there are
four sources of potential error.
1. The thermometer is not properly calibrated.
2. The thermometer needs a stem correction.
3. The thermometer is improperly located.
4. In distillations and boiling point determinations, the
effect of ambient pressure on the vapor pressure and consequently the observed
boiling point can be significant.
In the past, we used
mercury thermometers where the error was so small (except for case 3 in distillations!) that we could be sloppy and
ignore it. The measured temperatures
were reasonably close for our purposes.
Mercury is toxic and an environmental hazard. Over time thermometers broke and even with the best efforts, the
mercury could not be completely recovered.
Fortunately, new thermometers are available that use organic oils with a
small amount of a red or green dye. The
oils are much safer but at a price.
Mercury was chosen because for various reasons, it provides the most
accurate and reproducible temperature readings. Nothing else behaves quite as well for standard thermometers.

Calibration should not be a problem
for any thermometer. It wasn't in the
past for mercury and shouldn't be for the oils. Some of the new thermometers were checked. Although they are accurate at and above room
temperature, they might have a problem around 0° C. When the time
comes to measure a temperature below 10° C, you should check your thermometer to see if it is
properly calibrated. Your thermometers
are the type known as stem or partial immersion thermometers. You will notice that nearly 2 cm below the
-10°
C line, there is a thin brown ring.
Partial immersion thermometers are designed to give accurate readings if
they are dipped (immersed) into a liquid up to that point. To check your thermometer at lower
temperatures, suspend it in a stirring mixture of ice and water (an ice bath)
up to the immersion ring. After five
minutes, check the temperature. If it
reads 0°
C, it is properly calibrated, and you don't need to worry. If it reads 2° C for example, then you should subtract 2° C from any
temperature reading below 10° C. If your
thermometer reads for another liquid 8° C, the temperature is actually 6° C. You will also need to check at 100° C. Heat a beaker of water to boiling and
suspend the thermometer to its immersion for 5 minutes. If it does not read 100° C, then you will
have to adjust all future temperatures above 90° C accordingly.
Stem corrections are needed most
often when measuring vapor temperature in a distillation. Usually it's not needed for mercury partial
immersion thermometers. If you look at
the diagram of the proper location of a thermometer in a distillation head, you
can see the problem. The portion of the
thermometer suspended in the hot vapor is at a different temperature from the
remainder of the thermometer at the cooler room temperature. The mercury will not expand in the cool stem
to the same extent as the portion in hot vapor. Perhaps more importantly for immersion thermometers, they were
calibrated at 40°
C and not room temperature and this must be taken into account.
The equation for a
partial immersion thermometer to correct for this would be
Stem correction = a(DT)(T1 - T2)
a
= coefficient of expansion for the liquid.
For Hg, it is 0.000154. For most
organic liquids, it is 0.001.
DT
= the temperature difference between the temperature reading of the thermometer
and where the stem emerges from the thermometer adaptor.
For example, the
temperature may read that the vapor is 136° C. The first line
on scale that you may read on thermometer sticking out of the adaptor may be 33° C. DT would then be 136 - 33 = 103° C.
T1 = the
temperature at which the thermometers were calibrated. The manufacturer told us that it was 40° C
T2 = room
temperature. It is best if the room temperature
is measured near the cool portion of the stem with another thermometer.
Notice that for
mercury, a
is an order of magnitude smaller than for organic liquids. When we plug this value into the equations,
the difference in temperature is small enough for many applications that we can
be sloppy and ignore it. For the
organic oils, corrections of several degrees are not unusual and perhaps we
can't be sloppy without affecting our experiment. It will probably be necessary to do this calculation whenever you
do a distillation. An example is shown
below.
An organic solvent,
cyclohexanone, was distilled. The vapor
temperature read 152° C. Room temperature
was 24°
C, and the stem emerged from the adaptor at 29° C. The correction
would be
X = 0.001(152 -
29)(40 - 24) = 2°
C.
The actual vapor
temperature would be 152 + 2 = 154° C.
Since errors were
small for the mercury thermometers, then placement of the thermometer was not
as crucial. For the oil thermometers,
it will be important particularly for distillation.
Finally, the actual
boiling point will depend on the pressure of the system (barometric pressure
unless it is a vacuum distillation).
The Clausius-Clayperon equation can be used to estimate the boiling
point that should be observed. Unless
told otherwise, we won't need to worry about it.